FujiFilm DR-ID 1300 Service Manual page 325

Digital radiography
Table of Contents

Advertisement

l MODE 2 EXPOSURE TIME
Sets the charge accumulation time of READ MODE 2. Set in a range of 1.2 to 3.8 s.
The setting can be changed in 0.2-s steps.
<Settable range>
- 1.2 to 3.8 s [default: 1.2 s]
Charge accumulation time of READ MODE 2
u NOTE u
For the facilities that may use the machine long time under high temperature, to
prevent artifact, it is recommended to set the charge accumulation time of READ
MODE 2 at or lower than 1.2 seconds.
AUTO DETECT
l
Set whether to use the automatic X-ray detection function (AUTO DETECT) or not.
<Options>
- OFF [default]:
Not used
- ON:
Used
l CONNECTING CU
Set to CONNECTED when the machine is to be connected to the DX Console to
which AcSelarete or CALNEO U/T is connected.
Usually, use the setting of NOT CONNECTED [default].
<Options>
- NOT CONNECTED [default]
- CONNECTED
l EMERGENCY MODE
Set whether the emergency mode is to be used or not.
<Options>
- OFF [default]:
Not used
- ON:
Used
REFERENCE
Emergency mode is the mode that the calibration would be stopped automatically and
the exposure would be ready when the system starts.
When it is set to ON, the exposure would be ready as soon as the system starts.
018-234-03E
l BATTERY WARNING[min]
Sets the threshold value of remaining battery charge.
If the remaining battery charge lowers to the set threshold value, exposure cannot be
accepted. The SE LED flashes in this case.
The setting is made as the remaining battery time (minutes).
Always use the setting of 10 [default] in this machine.
<Settable range>
- 0 to 120 minutes [default: 10]
l SE BATTERY WATCHING (when connecting cables)
Set to ON when the MP does not exist in the facility.
<Options>
- OFF [default]:
- ON:
l MC-PC EXISTENCE
Set the availability of the PC for the MC.
<Options>
- NOT EXIST:
- EXIST [default]:
l A SYNCHRONOUS MC-SHUTDOWN WITH CONSOLE
Set the termination processing queue control of the MC and the Console.
<Options>
- SYNC:
- ASYNC [default]:
l LONG TIME ACCUM MODE
Set whether long time accumlation mode is to be used or not by selecting "ON" for
"AUTO DETECT" and "OFF" for "EMERGENCY MODE.
<Options>
- OFF [default]:
- ON:
REFERENCE
- It takes 10 seconds to read the images by setting "ON" to "LONG TIME ACCUM
MODE" only when the specified menus are selected in the DX Console. When the
common menus are selected, the reading time is the one that set by "READ MODE".
- When setting "ON" to "LONG TIME ACCUM MODE", set 7 minutes or more to
"SLEEP START TIME" in "SE SLEEP MODE SETTING" window.
DR-ID 1300 / DR-ID 1300PU Service Manual
Not watched
Watched
Select this option when no PC is available for the MC.
Select this option when the PC is available for the MC.
Select this option when no PC is available for the MC.
Select this option when the PC is available for the MC.
Not used
Used
MU2-60