n Mounting the AC Bucky Relay Unit (Optional)
NOTE
u
u
Abut the cable terminals as shown below when fastening them together.
GOOD
Cable
Screw
Terminal
Terminal block
NO GOOD
(1) Mount the AC bucky relay unit (optional) on the MP.
#1 Tighten: TP3x6 (x4)
AC bucky relay unit
MP
018-234-02E
600_700132.ai
#1
1200_700147.ai
(2) Connect the cable terminals with the terminal block of the MPX54A
board.
#1 Connect: Cable terminals
MPX54A Board
MPX2
MPX2-1
MPX2-2
MPX2-3
MPX2-4
MPX2-5-1
MPX2-5-2
MPX2-10
MPX2-11
*: Short cable (supplied accessory)
DR-ID 1300 / DR-ID 1300PU Service Manual
(*)
(*)
AC bucky relay unit
MPX1
MPX1-1
KA3
MPX1-2
KA4
MPX1-3
KB3
MPX1-4
KB4
MPX1-5-1
MPX1-5-2
KC5
KC1
KD3
KD4
MPX1-10
KE3
MPX1-11
KE4
KF5
KF1
IN2-29
600_700036.ai
IN2-29