FujiFilm DR-ID 1300 Service Manual page 546

Digital radiography
Table of Contents

Advertisement

(7) Check that the installation is completed, and click [Install SE MCU
Software].
#1 Check: OK
#2 Click: [Install SE MCU Software]
Install SE1 Software>>window
#1
NOTE
When the SE MCU software is updated successfully, "Successed Please Reboot
SE OK" appears after about 80 seconds. In this case, this procedure is completed
successfully. Updating the SE application software version is completed after
restarting the SE.
In addition, if about 40 seconds elapse after "OK" appears, "Make sure if MCU
software is latest version." appears. Ignore this message if "Successed" has
already appeared.
If "Successed" has not appeared and only "Make sure if MCU software is latest
version." is displayed, restart the SE and check the SE application software
version.
If the update is not performed, update the SE MCU software version once again.
(8) For the other panel units, perform steps (2) to (7) as required.
018-234-04E
#2
1200_700123E.ai
(9) Check that the installation is completed, and restart the MP.
#1 Check: OK
Install SE1 Software>>window
#1
(10) Check the SE application software version.
{IN2:10.10.1_Checking the SE Application Software Version}
DR-ID 1300 / DR-ID 1300PU Service Manual
IN2-69
1200_700124E.ai
IN2-69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dr-id 1300pu

Table of Contents