(5) Check for sensitivity and density problems.
Check to make sure that the system sensitivity value and the film density value
appearing on the output film are approximately 200 and 1.2, respectively.
u NOTE u
If the S value is more than 300 or less than 100, the Machine-Specific Data may
not be correctly installed. In such a case, reinstall the Machine-Specific Data and
perform "11. Calibration".
{IN2:10.12_Installing the Machine-Specific Data}
REFERENCE
If the gain calibration is not performed with SID = 215 cm or more, the S value of
the TOP/BOTTOM panel unit may deviate from the S value of the CENTER panel
unit.
018-234-04E
INSTRUCTION
If the images have any problems, perform "11.1 Offset Calibration", "11.2 Gain
Calibration", "Appendix 5.1 Defect Calibration", and "Appendix 5.2 Lag Calibration" in
this order, and then perform "13.1 Checking for Irregularities, Density Problems, White
Blank Portion, and Sensitivity Problems in the Images" again. If the images of the
CENTER panel unit have problems, "Appendix 5.1 Defect Calibration" is not required
to be performed.
{IN2:11._Image Calibration}
{IN:Appendix 5._Defect Calibration and Lag Calibration}
If the problems still remain in the images after the countermeasures above, obtain the
data below and request troubleshooting.
- The capture image of the DX Console screen
Indicate abnormal images and portions of the images by using circles and arrows.
- Information on the facility and miscellaneous
- Frequency
- Repeatability
- Whether the problem occurs with other panels.
- Exposure conditions (kV, mA, msec, SID, grid, irradiation fields)
- Software version (MC and DX Console)
- Something you found
- Log (LOG ALL)
Secure and collect corresponding logs as soon as quickly before carelessly
eliminating them. Immediately after the problem occurs, check that the log includes
the time of the problem (the log has been updated).
- Calibration data (CORRECT)
Even if, for example, only the TOP panel is abnormal, the CENTER and/or the
BOTTOM panel units may relate in part to the problem. Collect all calibration data
for the three panel units.
- Input images to the MC software (a total of three images each for TOP, CENTER
and BOTTOM)
In the case where the raw image saving mode is OFF, the latest 50 exposures of
images before the stitching area correction are saved by the procedures below.
Collect corresponding one exposure (three images).
1. Use the LOG ALL of the PC TOOL to save the log.
A "LP_IMAGE" folder is newly created in the folder saved with the LOG ALL.
2. Perform "BackupLPImage.bat" in the "LP IMAGE" folder.
The latest 50 exposures of raw images are saved in the "LP IMAGE" folder.
- Routine files (a total of four images each for TOP, CENTER, BOTTOM and
Synthesis)
Obtain complete set of the Routine files having extensions other than "inf" that
includes personal information.
DR-ID 1300 / DR-ID 1300PU Service Manual
IN2-87
IN2-87