10.13.2 Checking the SE Correct Data Save Status
(1) Click [SE Correct Data Save Status >>] and input the target SE
number, then click [SET].
#1 Click: [SE Correct Data Save Status >>]
#2 Input: Target SE number
#3 Click: [SET]
CALIBRATION window
018-234-04E
#1
SE Correct Data Save Status >>
window
#2
#3
1200_700186E.ai
(2) Check that saving of the SE correct data is completed.
#1 Check: Save result
SE Correct Data Save Status >> window
u INSTRUCTION u
Mode3 is not used for DR-ID 1300 and thus "NG" is displayed, but you can ignore
this judgment.
(3) For other two panel units, perform steps (1) and (2).
DR-ID 1300 / DR-ID 1300PU Service Manual
IN2-74
#1
1200_700187E.ai