FujiFilm DR-ID 1300 Service Manual page 89

Digital radiography
Table of Contents

Advertisement

Error
Error Name
Code
Exposure menu registration
11247
error
Mode 2 accumulation time
11248
mismatch error
11271
MCU error
11404
DAP communication error
X-ray irradiation before it is
11408
ready for exposure error
11602
Correction data load error
Correction data transfer error
11603
(MC → SE)
MP communication connection
11750
error
11751
MPC fuse blown error
Panel connector disconnection
11752
in MP
11900
Parameter error
Memory map reconfiguration
12150
retry
018-234-06E
Occurrence Condition
[During registration of exposure menu]
An error occurs in exposure menu registration.
[During bootup]
Values of the mode 2 accumulation time set in the SE and the MC do not
match.
The SE detects an error of MCU (microcomputer) and informs the MC.
[During bootup/routine processing]
A DAP communication error occurs.
[During routine processing]
At the time of active AUTO DETECT, exposure has been implemented
after exposure menu registration before completion of the exposure
preparation (before lighting of the ShotReady icon of the CL).
[During bootup]
Read of the correction data load file fails (when the maximum retry count
is exceeded).
<Reference>
You can check the file name you failed to read on the error log.
[During bootup]
An error occurs in transfer of the correction data from the MC to the SE.
[During bootup/routine processing]
Disconnection of a line from the MP is detected.
[During bootup/routine processing]
A blown fuse on the MPC board is detected.
During bootup/routine processing]
Disconnection of the panel connector connected in the MP is detected.
[During bootup]
An error is detected in an item set by service personnel (wrong value
setting, inconsistency of combination, or the like)
[During bootup/routine processing]
A retry of memory map reconfiguration fails.
DR-ID 1300 / DR-ID 1300PU Service Manual
Probable Cause and Remedy
- Check the connection with the AP (for wireless type).
- Check the connection of the SE cables (for wired type).
- Start up the MUTL and set the mode 2 accumulation
time.
- The SE detects an error of the microcomputer.
Check the DAP relay cable.
- Register the exposure menu again and expose again.
- Rewrite the file you failed to read.
- Check communication among the MC, the MP and the
FPD.
- Check the connection with the AP (for wireless type).
- Check the connection of the SE cables (for wired type).