FujiFilm DR-ID 1300 Service Manual page 501

Digital radiography
Table of Contents

Advertisement

(5) Connect the cable terminals with the terminal block of the MPX54A
board.
#1 Connect: Cable terminals
#1
MPX2
MPX1
1 5V2 ISO
2 SHOT2 RDY1
3 SHOT2 RDY1
4 SHOT2 RDY2
5 SHOT2 RDY2
6 SHOT2 GO A
7 SHOT2 GO B
8 BKY2 RDY1A
9 BKY2 RDY1B
10 BKY2 RDY2A
11 BKY2 RDY2B
12 NC2
#1
018-234-02E
X-Ray shot cable
1 5V1 ISO
2 SHOT1 RDY1
3 SHOT1 RDY1
4 SHOT1 RDY2
5 SHOT1 RDY2
6 SHOT1 GO A
7 SHOT1 GO B
8 BKY1 RDY1A
9 BKY1 RDY1B
10 BKY1 RDY2A
11 BKY1 RDY2B
12 NC1
Bucky cable
600_700028.ai
(6) Wrap a plastic tape around the unconnected (remaining) cable
terminals.
u NOTE u
No cable terminal may remain depending on the X-ray high voltage generator to
be connected in some cases. If no cable terminal remains, the procedure is not
necessary.
#1 Wrap: Plastic tape
DR-ID 1300 / DR-ID 1300PU Service Manual
#1
600_700014.ai
IN2-24
IN2-24