FujiFilm DR-ID 1300 Service Manual page 206

Digital radiography
Table of Contents

Advertisement

 SE Correct Data Save Status
After finishing the full calibration or the SE correct data transfer, check whether the
correction data storage has been completed or not.
(1) Click [GET].
Depending upon the saving status, the following will be displayed. Verify that
transferring has completed.
"PROCESSING" "COMPLETED" "TRANSFER FAILED"
If the transferring failed, then re-implement a forced transfer of the correction
data.
(2) Confirm the correct data save status.
018-234-06E
 SE Correct Data Status (Memory)
(1)
(2)
1200_400077E.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
Check the list of correct data stored by the memory in the SE.
(1) Click [GET].
The list of correct data is displayed .
(2) Confirm the list of correct data.
MU1-41
(1)
(2)
1200_400078E.ai
MU1-41