REFERENCE
- In comparison with High-Speed Mode, the time required for image acquisition in the
Standard Mode is approximately 370 ms longer.
- Even if the noise is generated via image evaluations whereas physical phantoms
such as mesh, etc. were used, this noise is not generated in human body exposures.
IMAGE LOG SAVE
Selects whether or not the recorded image log is to be stored in the HDD.
Always use the setting of OFF [default] in the machine.
<Options>
- OFF [default] :
Not stored
- ON:
Stored
u NOTE u
When set to "ON", the PC capacity is compressed and the throughput is degraded.
IMAGE LOG SAVE TO FTP
Selects whether or not the recorded image log is to be stored in the FTP.
Always use the setting of OFF [default] in the machine.
<Options>
- OFF [default] :
Not stored
- ON:
Stored
u NOTE u
When set to "ON", the PC capacity is compressed and the throughput is degraded.
OFFSET2 (blemish/unevenness correction setting)
Select whether or not the blemish/unevenness correction is to be performed.
<Options>
- OFF:
Not performed
- ON [default]:
Performed
T2 TIMER(sec)
Set the start time (sec) of the calibration executed in the background. When the
exposure has been completed, the calibration begins after the time elapsed that was
set.
- Character string: [default: 120]
018-234-06E
AUTO DEFECT CALIBRATION
Set whether the automatic defect calibration is to be executed or not.
<Options>
- OFF:
- ON [default]:
AUTO DEFECT CALIBRATION START HOURS
When "ON" is selected in "AUTO DEFECT CALIBRATION", input the automatic
execution time hourly.
- 2-character string or less [default: 4]
AUTO DEFECT CALIBRATION START MINUTES
When "ON" is selected in "AUTO DEFECT CALIBRATION", input the automatic
execution time minutely.
- 2-character string or less [default: 0]
OFFSET JUDGING
Select whether or not the blemish/unevenness correction is to be performed.
<Options>
- OFF:
- ON [default]:
DR-ID 1300 / DR-ID 1300PU Service Manual
Not executed
Executed
Not performed
Performed
MU1-19
MU1-19