FujiFilm DR-ID 1300 Service Manual page 556

Digital radiography
Table of Contents

Advertisement

(4) Click [Start].
"OK" appears in the result display area upon completion of the preparation for
exposure.
u NOTE u
If you click [Start] during automatic offset update (for 30 seconds every approx.
10 minutes), "Error 12700 currently unavailable" appears. Click [Start] again after
an interval of approx. 30 seconds, in this case.
(5) Click [Display ShotRdy] and confirm display of "SHOT_READY" in the
result indication column.
u NOTE u
If "NOT_READY" is indicated, wait for approx. 5 seconds, try again, and proceed
to the next step (exposure) after "SHOT_READY" is indicated.
When exposure is executed in "NOT_READY" status, the calibration fails and an
abnormal image is generated.
(6) Expose 16 times with the following condition. Click [Display ShotRdy]
at each exposure and confirm display of "SHOT_READY" in the result
indication column.
REFERENCE
Clicking [Display Image Count], you can know the number of exposures.
The count is displayed in the format of "n/16" in the result display area after
n-times of exposures.
- 1305SE : Tube voltage of 75 kV, Dose of 10 mR
Images for gain correction data generation (16 frames) are read from the SE (FPD).
u INSTRUCTION u
Set the exposure time to 200 msec or less. Exposure for longer than 200 msec
cannot be made, since the maximum exposure time is inherently specified as
200 msec.
REFERENCE
Exposure condition examples for exposing an IP to a dose of 10 mR are shown
below. (1305SE)
SID:
215 cm
Voltage:
75 kV
Amperage: 100 mA
Time:
90 msec
018-234-04E
(7) Click [Create a Gain Data].
The confirmation dialogue box of "Are you sure?" opens.
REFERENCE
If you click [Create a Gain Data] when less than sixteen exposures have been
made, "Error!!" appears in the result display area. Continue to make exposures
until sixteen exposures are made, and click [Create a Gain Data], in this case.
(8) Click [OK].
When the defect correction data is generated and stored in the HDD, [OK]
appears in the result display area.
(9) Click [BACK] to return to the Calibration window.
NOTE
If the calibration freezes in the above procedure (5) or (6), left-click the MC Manager
from the task tray and execute "EXIT". Check that the MC Manager icon has
disappeared from the task tray, and then start the MC Manager from "Start menu" →
"Start-up".
(10) Select "Shut Down" in the menu of the DX Console.
(11) Check that the READY LED (green) of the SE flashes and becomes
unlit, and the calibration has been completed when the system
terminates. Then, wait for the DX Console being powered OFF.
(12) Turn OFF the power of the SE.
(13) Turn ON the power of the SE, then turn ON the power of the DX
Console.
DR-ID 1300 / DR-ID 1300PU Service Manual
IN2-79
IN2-79