FujiFilm DR-ID 1300 Service Manual page 627

Digital radiography
Table of Contents

Advertisement

- Optional item
 SE cable
 10m
 20m
 X-Ray Shot Cable
 3-code / 5m
 3-code / 15m  9-code / 5m
 AC Bucky Relay Unit
 100-110V
 110-120V
 MP anchor retaining bracket
 I/F box  Hand switches (2 buttons)
 I/F box cable (Used to connect between the I/F box and the MP)
 10m
 15m
 I/F box cable (Used to connect between the I/F box and the X-ray equipment)
 110-3 wire
 110-4 wire
 Shimadzu
 Siemens
 DelMedical
 DelMedical (round terminal)
 Philips
3. Checking for short-circuiting of the power supply voltage/machine
power source
{IN1:6._Installing the MP}
 Single-phase two-wire
 Single-phase three-wire
- Power supply voltage: __________V
Specified value (Japan): 100 VAC ±10%
Specified value (Outside Japan): 100-240 VAC ±10%
- Power cable:
Resistance value measured between E and FG:
Resistance value measured between N and FG: ______Ω
Resistance value measured between L and FG: ______Ω
Resistance value measured between L and N: ______Ω
018-234-04E
 N/A  OK  NG
 N/A  OK  NG
 9-code / 15m
 N/A  OK  NG
 200-220V
 220-240V
 N/A  OK  NG
 N/A  OK  NG
 N/A  OK  NG
 GE1
 GE2
 Toshiba
 CPI
 GENDEX
(specified value: with continuity)
(specified value: ∞)
(specified value: ∞)
(specified value: 100 kΩ or higher)
4. Checking the SE
TOP SE panel serial ID:
CENTER SE panel serial ID:
BOTTOM SE panel serial ID:
4.1 Checking connection with X-ray high voltage generator
{IN1:6.5_Connecting the X-Ray Shot Cable}
- Manufacturer of the connected
X-ray high voltage generator:
- Model:
- Tube No.:
 Exposure stand Manufacturer of the incorporated : _______________________
 Grid:
 Steady grid type
4.2 Terminal block having the X-Ray Shot Cable connected
- MP [ ____ ]
 MPX0
4.3 Changing the local IP address
{IN1:10.13_Changing the IP Address of the SE/MP/MC and the FTP Server}
{IN2:10.14_Changing the IP Address of the SE/MP/MC}
 OK  NG
- MP IP address:
- TOP SE panel IP address:
- CENTER SE panel IP address:
- BOTTOM SE panel IP address:
4.4 Use of AUTO DETECT
 OK  NG
 No
 OK  NG
4.5 Calibration
 OK  NG
{IN1:11._Image Calibration}
{IN2:11._Image Calibration}
 OK  NG
4.6 Marker Calibration
{IN1:12._Marker Calibration}
{IN2:12._Marker Calibration}
DR-ID 1300 / DR-ID 1300PU Service Manual
___________________
___________________
___________________
_______________________________
_______________________________
_______________________________
Model: _______________________
 Bucky contact type
 MPX1
_______________________________
_______________________________
_______________________________
_______________________________
Yes
PC-2
OK
NG
 Bucky AC type
OK
NG
 N/A  OK  NG
 N/A  OK  NG
 OK  NG
 OK  NG
PC-2