- Optional item
SE cable
10m
20m
X-Ray Shot Cable
3-code / 5m
3-code / 15m 9-code / 5m
AC Bucky Relay Unit
100-110V
110-120V
MP anchor retaining bracket
I/F box Hand switches (2 buttons)
I/F box cable (Used to connect between the I/F box and the MP)
10m
15m
I/F box cable (Used to connect between the I/F box and the X-ray equipment)
110-3 wire
110-4 wire
Shimadzu
Siemens
DelMedical
DelMedical (round terminal)
Philips
3. Checking for short-circuiting of the power supply voltage/machine
power source
{IN1:6._Installing the MP}
Single-phase two-wire
Single-phase three-wire
- Power supply voltage: __________V
Specified value (Japan): 100 VAC ±10%
Specified value (Outside Japan): 100-240 VAC ±10%
- Power cable:
Resistance value measured between E and FG:
Resistance value measured between N and FG: ______Ω
Resistance value measured between L and FG: ______Ω
Resistance value measured between L and N: ______Ω
018-234-04E
N/A OK NG
N/A OK NG
9-code / 15m
N/A OK NG
200-220V
220-240V
N/A OK NG
N/A OK NG
N/A OK NG
GE1
GE2
Toshiba
CPI
GENDEX
(specified value: with continuity)
(specified value: ∞)
(specified value: ∞)
(specified value: 100 kΩ or higher)
4. Checking the SE
TOP SE panel serial ID:
CENTER SE panel serial ID:
BOTTOM SE panel serial ID:
4.1 Checking connection with X-ray high voltage generator
{IN1:6.5_Connecting the X-Ray Shot Cable}
- Manufacturer of the connected
X-ray high voltage generator:
- Model:
- Tube No.:
Exposure stand Manufacturer of the incorporated : _______________________
Grid:
Steady grid type
4.2 Terminal block having the X-Ray Shot Cable connected
- MP [ ____ ]
MPX0
4.3 Changing the local IP address
{IN1:10.13_Changing the IP Address of the SE/MP/MC and the FTP Server}
{IN2:10.14_Changing the IP Address of the SE/MP/MC}
OK NG
- MP IP address:
- TOP SE panel IP address:
- CENTER SE panel IP address:
- BOTTOM SE panel IP address:
4.4 Use of AUTO DETECT
OK NG
No
OK NG
4.5 Calibration
OK NG
{IN1:11._Image Calibration}
{IN2:11._Image Calibration}
OK NG
4.6 Marker Calibration
{IN1:12._Marker Calibration}
{IN2:12._Marker Calibration}
DR-ID 1300 / DR-ID 1300PU Service Manual
___________________
___________________
___________________
_______________________________
_______________________________
_______________________________
Model: _______________________
Bucky contact type
MPX1
_______________________________
_______________________________
_______________________________
_______________________________
Yes
PC-2
OK
NG
Bucky AC type
OK
NG
N/A OK NG
N/A OK NG
OK NG
OK NG
PC-2