FujiFilm DR-ID 1300 Service Manual page 492

Digital radiography
Table of Contents

Advertisement

n Signal Descriptions
l SHOT1 RDY1/RDY1 (Input)
The signal is generated when only the first stage of the shot switch on the X-ray high
voltage generator is pressed. (Prep signal)
l SHOT1 RDY2/RDY2 (Input)
The signal is generated when the first and second stages of the shot switch on the
X-ray high voltage generator are pressed. (RAD signal)
l SHOT1 GO A/B (Output)
The signal generates X-rays.
A relay type is normally open (NO).
n Contact Capacity
The specifications of relays (K4, K5 and K6) are as follows.
l Relay model
AHN22005 (Panasonic Electric Works)
l Rated control capacity
Standard contact: 30 VDC
l Maximum switching current
1 A
018-234-02E
n Timing Chart
A: Preparation starts for SE exposures corresponding to the first stage of the switch when
the shot switch (first stage) is turned ON.
B: Preparation starts for SE exposures corresponding to the second stage of the switch
when the shot switch (second stage) is turned ON.
C: Charge accumulation of the SE (FPD) starts.
D: An exposure request signal is output, and X-ray exposure starts.
E: X-ray exposure ends.
F: Charge accumulation of the SE (FPD) ends.
SHOT1 RDY1/RDY1
SHOT1 RDY2/RDY2
SHOT GO
X-ray exposure
Charge accumulation
of SE (FPD)
DR-ID 1300 / DR-ID 1300PU Service Manual
A
B
C
D
E
F
600_700011.ai
IN2-15