FujiFilm DR-ID 1300 Service Manual page 211

Digital radiography
Table of Contents

Advertisement

 Line defects registration
(1) Select "Register line defect data" from the DR Maintenance Software
screen's menu on the left.
(2) Click the pull-down menu in "Select target SE", and select the target
SE.
u NOTE u
The SE (FPD) that can be selected in this menu is the SE which is registered in
the SE registration. Note that it is not the "SE which is connected".
(3) In the "Enter the vertical line defect X coordinate", or the "Enter the
horizontal line defect Y coordinate" text box, enter the line defect
coordinates which were confirmed via ImageJ.
REFERENCE
Up to five vertical line defects, and up to five horizontal line defects can be
simultaneously registered.
(4) Click [Set].
018-234-06E
If data registration is successful, the following screen is displayed.
REFERENCE
If it fails, recheck the FPD connections.
If an overabundance of line defect data are registered is displayed, no more
correction data can be registered.
(5) Restart the PC.
(6) Confirm that the PC and CSL are started up, and that the target FPD is
being recognized by the CSL.
(7) Select "Complete System" from CSL.
Calibrating will be executed when it is shut down.
(8) Start the PC again.
(9) Start the DR Maintenance Software.
(10) Select the "OPERATION" - "CALIBRATION" - "SE Correct Data
Transfer".
(2)
(11) Select the target FPD and execute "SE Correct Data Forced Transfer".
{MU1:1.4.5_SE Correct Data Transfer}
(3)
(4)
1200_700567.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
MU1-46
1200_700568.ai
MU1-46