4.1.2
Further Information for IEC 60601-1-2 (EN60601-1-2)
- Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
the accompanying documents.
- Portable and mobile RF communications equipment can affect medical electrical
equipment.
- Information regarding the cable affecting EMC is as follows.
Name
Connected Device
Power cable
DR-ID 1300PU
- The use of accessories, transducers and cables other than those specified, with the
exception of transducers and cables sold by FUJIFILM Corporation as replacement
parts for internal components, may result in increased emissions or decreased
immunity of the DR-ID 1300PU.
- The DR-ID 1300PU should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the DR-ID 1300PU should be observed to
verify normal operation in the configuration in which it will be used.
- Basic performance of the equipment and the system
Image data are acquired from the DR-ID 1300PU and are saved in and displayed on
the DR-ID 300CL.
018-234-02E
Maximum
General Specification
Length
Use a hospital-grade power cable.
(for North America)
3 m
A non-hospital grade power cable
can be used.(for other countries)
Guidance and manufacturer's declaration - electromagnetic emissions
The DR-ID 1300PU is intended for use in the electromagnetic environment specified below.
The customer or the user of the DR-ID 1300PU should assure that they are used in such
an environment.
Emissions test
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class A
Harmonic emissions
Complies
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
* IEC61000-3-2, which set limits for harmonic emissions, and IEC61000-3-3, which set limits
for voltage fluctuations/flicker emissions, are applicable to equipment with power-supply
voltage over 220 V. It is not necessary to apply these standards to equipment with power-
supply voltage less than 220 V.
DR-ID 1300 / DR-ID 1300PU Service Manual
Safety Precaution
Electromagnetic
Compliance
environment - guidance
The DR-ID 1300PU uses
RF energy only for their
internal function.
Therefore, their RF
emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
The DR-ID 1300PU is
suitable for use in all
establishments other
than domestic and those
directly connected to
the public low-voltage
power supply network that
supplies buildings used for
domestic purposes.
Safety Precaution