(3) Move the irradiation field in the opposite direction as shown below.
1300_700033.ai
(4) After checking that the green [Ready] button is lit, expose the 2nd
shot with the following conditions. Confirm that [Ready] lights with
the green button at each exposure.
REFERENCE
Perform the next exposure every time after [Ready] is lit up with the green button
because the [Ready] disappears every exposure.
- Tube voltage of 80 kV, Amperage: 100 mA, Time: 20 msec, SID: 120 cm
018-234-05E
(5) After checking that the green [Ready] button is lit, expose the 3rd shot
with the following conditions. Confirm that [Ready] lights with the
green button at each exposure.
REFERENCE
Perform the next exposure every time after [Ready] is lit up with the green button
because the [Ready] disappears every exposure.
- Tube voltage of 80 kV, Amperage: 100 mA, Time: 20 msec, SID: 120 cm
u NOTE u
Do not move the irradiation field.
1300_700033.ai
(6) Return the irradiation field to the position of the first shot as shown
below.
1300_700034.ai
(7) After checking that the green "Ready" button is lit, expose the 4th
shot with the following conditions. Check that the exposure count
turned to "4/4" .
- Tube voltage of 80 kV, Amperage: 100 mA, Time: 20 msec, SID: 120 cm
DR-ID 1300 / DR-ID 1300PU Service Manual
IN1-83
IN1-83