 Operation check
(1) Restart the FPD.
At this time, no errors should occur in the FPD initialization.
(2) Do not place any subjects on the FPD, and select the "Image format"
exposure menu, and irradiate the X-rays.
At this time, there are not any image anomalies.
(3) Back up the correction data for the target FPD with the DR
maintenance software.
(4) Check that the registered coordinates are written in the log files in the
backed up correction data.
If there are vertical line defects, refer to DV.txt, and for horizontal line defects refer to
DH.txt.
u INSTRUCTION u
If the line defects are registered in incorrect locations, restore the correction data
which was obtained in "■ Preparation (1)", and re-execute.
REFERENCE
If the line defects with 3 or more lines continue in both the vertical and horizontal
directions, or if the intervals between the line defects are 3 lines or less, a 12231
error will occur.