FujiFilm DR-ID 1300 Service Manual page 300

Digital radiography
Table of Contents

Advertisement

l Battery:
u NOTE u
Not used in the DR-ID 1300.
Displays the following information on the battery in the result display area.
Serial No.: 0 to 65535
Date of manufacture: YYYY/MM/DD
Battery Status: 0xXXXX
Failure cause (Manufacture Access): 0bXXXXXXXXXXXXXXXX
Battery capacity: 0 to 65535 [unit mAh/10 mWh]
Fitting (putting on/off) count of battery (B/I Logic Count): 0 to 65535
Charge/discharge count (Cycle Count): 0 to 65535
Remaining time (Time to Empty): 0 to 65535 [unit min]
Battery temperature: 0 to 65535 [0.1 K]
<Display example>
Serial No = 66
Manufacture Date = 2010/09/23
Battery Status = 0x0080
Manufacture Access = 0b1110000000000000
Battery Capacity: 2619 (unit mAh/10 mWh)
B/I Logic Count: 0
Cycle Count: 27
Time to Empty: 65535 (min)
Temperature: 2956 (0.1 K)
l Battery Date Setting
u NOTE u
Not used in the DR-ID 1300.
Set the start date for battery use.
Accurate display of the battery capacity becomes possible by performing "Battery
Date Setting".
018-234-02E
l GLG Check
u NOTE u
Not used in the DR-ID 1300.
The following tests are performed for the GLG board, and the results (OK/NG) are
displayed in the result display area.
- Memory check (writing of data to and reading of data from FRAM and RAM)
- Data acquisition from RTC
- Low G sensor data setting and data acquisition
- AD converter diagnosis
- Reading of set data
REFERENCE
Self-diagnosis of the GLG board is performed also at the time of equipment start.
DR-ID 1300 / DR-ID 1300PU Service Manual
MU2-35
MU2-35