<"RU IMAGE FLAG" window>
RU IMAGE FLAG window
l GOS SHARPNESS FILTER
Select the sharpness of the GOS panel.
Select HIGH when a highly sharp image is to be obtained, making use of the features
of DR.
Select NORMAL when an image with normal sharpness close to the FCR is to be
obtained.
<Options>
- HIGH [default]:
Highly sharp
- NORMAL:
Normally sharp
u NOTE u
As this setting causes a large change of the image quality, select according to the
desire of the user.
018-234-03E
1200_400037.ai
l IMAGE READING MODE
Selects the image compensation processing mode.
Normally use the default setting. If the noise is generated via image evaluations
whereas physical phantoms such as mesh, etc. were used, change to the Standard
Mode.
<Options>
- Standard Mode
- High-Speed Mode [default]
u NOTE u
When this setting has been changed, perform the full calibration after restarting the
MC in the following procedure. If not, the image compensation processing mode is not
switched correctly.
1. Change the setting.
2. Restart the MC.
3. Perform the full calibration.
REFERENCE
- In comparison with High-Speed Mode, the time required for image acquisition in the
Standard Mode is approximately 370 ms longer.
- Even if the noise is generated via image evaluations whereas physical phantoms
such as mesh, etc. were used, this noise is not generated in human body exposures.
l IMAGE LOG SAVE
Selects whether or not the recorded image log is to be stored in the HDD.
Always use the setting of OFF [default] in the machine.
<Options>
- OFF [default] :
- ON:
u NOTE u
When set to "ON", the PC capacity is compressed and the throughput is degraded.
DR-ID 1300 / DR-ID 1300PU Service Manual
Not stored
Stored
MU2-63