FujiFilm DR-ID 1300 Service Manual page 86

Digital radiography
Table of Contents

Advertisement

Error
Error Name
Code
RMV board communication
11112
error
11140
CPU image correction error
Exposure time sequence
11160
timeout
11201
FAN stop error
11210
FPD low temperature warning
11211
FPD high temperature warning
Battery high temperature
11215
warning
018-234-06E
Occurrence Condition
[During bootup/routine processing]
Although an attempt is made to set the reading mode for the RMV
board mounted on the FPD, an error in communication is detected when
reading back the register value after setting of the reading mode. The
communication is retried, but results in a failure. The FPD is disabled.
[During M-Utility]
- An error is detected by Communication Check of RMV.
[During routine processing]
Processing of the CPU image correction library fails.
[During routine processing]
Signals necessary for continuing the exposure cannot be received after
starting the exposure, resulting in timeout.
[During bootup/routine processing]
MP cooling fan stop is detected.
<Reference>
You can know which fan stops on the error log.
[During bootup/routine processing]
A low temperature (Approx. 10°C or lower) in the FPD is detected, and
the FPD is disabled.
<Reference>
- You can know the No. of the FPD of interest and the temperature when
the error is detected on the error log.
- This temperature is an estimate since it fluctuates depending on the SE
type and the external temperatures.
[During bootup/routine processing]
The panel gets high temperature (Approx. 40°C or higher) and is disabled.
<Reference>
- You can know the No. of the FPD of interest and the temperature when
the error is detected on the error log.
- This temperature is an estimate since it fluctuates depending on the SE
type and the external temperatures.
[During bootup/routine processing]
The battery temperature rises to the warning level.
DR-ID 1300 / DR-ID 1300PU Service Manual
Probable Cause and Remedy
- Select a defective matter from the detail error log
indicated in ERRORDB.LOG, and check it.
{MT:1.3.1_11112 Error}
- Check the MP cooling fan (P-FAN).
- Check the environments at the installation site.
- Check the use conditions.
- Check the FPD.
- Check the environments at the installation site.
- Check the use conditions.
- Check the MP cooling fan (P-FAN).
- Check the FPD.
- Check the ambient temperature.
- Space exposures to lower the temperature in the FPD.