(4) Connect the jig cable <supplied accessory> to the connector of the
RMV board.
#1 Disconnect: Connector cable (RMV16)
#2 Connect: Jig cable <supplied accessory>
RMV16
#1#2
(5) Connect the SE cable and turn ON the power of the MP.
018-234-02E
(6) Check to make sure that two external positions of the status lamp (LED)
of the SE flash three times successively.
You can know that the IP address of the SE is changed to the default value if the
LED flashes three times successively.
#1 Check: LED
SE
1300_700021.ai
(7) Turn OFF the power of the MP and disconnect the SE cable.
(8) Disconnect the jig cable you connect in the procedure (4) and connect
the connector cable (RMV16).
(9) Put on the SE rear cover.
{MC:3.1.1_SE Rear Cover}
(10) Set the COUNTRY CODE in "WIRELESS SETTING" window of EDIT
CONFIGURATION.
{MU2:1.11_EDIT CONFIGURATION}
DR-ID 1300 / DR-ID 1300PU Service Manual
AppxIN2-4