PANEL SETTING

1.2.1

SE Registration/Deletion

Used to register the serial ID of the panel unit, or to cancel a registered panel unit.
 SE Registration
Used to register the serial ID of the panel unit.
(1) Click "PANEL SETTING" - "SE Registration/Deletion".
The SE Registration/Deletion window opens.
(2) Click [Add].
The Panel Registration window opens.
(3) Input the serial ID and click [OK].
The serial ID of the SE is registered.
018-234-04E
(1)
(2)
1200_400051E.ai
1200_400055E.ai
u NOTE u
- The first letter of the serial ID must be entered as a capital letter.
- For SE registration of the long panel, first register three panels as the FPD panels.
This is the same as the conventional SE panel.
After that, for the long panel registration, specify the panel unit consisting of the
three panels - TOP, CENTER, and BOTTOM - for the panels registered above. The
panel unit where TOP, CENTER and BOTTOM are specified functions as one long
panel.
REFERENCE
The serial ID of an SE is mentioned on the surface of the machine-specific data
CDROM.
DR-ID 1300 / DR-ID 1300PU Service Manual
MU1-5
MU1-5