(6) Return the irradiation field to the position of the first shot as shown
below.
1300_700034.ai
(7) After clicking [Display ShotRdy] and confirming the display of "SHOT_
READY" in the result indication column, expose the 4th shot with the
following conditions. Click [Display Image Count] and check that the
exposure count turned to "4/4" .
- Tube voltage of 80 kV, Amperage: 100 mA, Time: 20 msec, SID: 120 cm
018-234-04E
n Creating Calibration Data and Checking Result
(1) Click [Create a Marker Data].
The confirmation dialogue box of "Are you sure?" opens.
Marker Calibration >> window
(2) Click [OK].
When the marker correction data is generated and stored in the HDD, [OK]
appears in the result display area.
u NOTE u
If the data generation failed, an error number appears on the status screen.
In that case, check the position of the calibration phantom and the irradiation
field, and then perform again the "■ Exposing Calibration". If an error still occurs,
obtain the following data and report the failure.
- LogAll (IMO.log / Calib.log)
(3) Check that the calibration data have been generated.
REFERENCE
The data are stored in the following location. Check the time stamp of the file
generated.
- C:\Program Files\FujiFilm\MC\opt\config\machine\correct\PANEL2\
LPMarkerData.txt
DR-ID 1300 / DR-ID 1300PU Service Manual
IN2-84
1300_700035E.ai
IN2-84