(5) Enter the IP address of the currently operating PC (192.168.0.10) and
press the <Enter> key.
The OS version update starts.
It takes up to approx. 5 minutes to complete the installation.
CAUTION
While write into the HDD is in progress, never turn OFF the MP power. If turned
OFF, the data on the HDD gets damaged, and the MP cannot boot up as a result.
018-234-06E
1200_300169.ai
1200_300170.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
(6) The MP restarts automatically after the following window appears.
(7) Press the <Enter> key to close the Command Prompt window.
u INSTRUCTION u
For two MP configuration, perform steps (2) to (7) for the other MP.
(8) Remove the MP OS Installer Ver. 1.0 media from the DVD drive for the
DX Console and restart the system.
(9) Check the MP OS version.
{MC: Appendix 1.1_Checking the MP OS Version}
AppxMC1-3
1200_300171.ai
AppxMC1-3