FujiFilm DR-ID 1300 Service Manual page 229

Digital radiography
Table of Contents

Advertisement

 Battery Start Date Setting
u NOTE u
Not used in the DR-ID 1300.
Set the start date for battery use.
Accurate display of the battery capacity becomes possible by performing "Battery
Start Date Setting".
(1) Click "CHECK DEVICE STATE" - "Check Board Status".
The Check Board Status window opens.
(2) Select the target SE in "Check SE Board Status"-"Battery Start Date
Setting" from the drop-down list box.
(3) Click [SET].
If the battery usage start day has been set, and it is successful, a pop up window
displaying "Succeeded" will pop up.
(4) Click [OK].
018-234-06E
 GLG Board Self-diagnostic
(1)
(2)
(3)
1200_400086E.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
u NOTE u
Not used in the DR-ID 1300.
The following tests are performed for the GLG board, and the results (Succeeded/
Failed) are displayed in the result display area.
- Memory check (writing of data to and reading of data from FRAM and RAM)
- Data acquisition from RTC
- Low G sensor data setting and data acquisition
- AD converter diagnosis
- Reading of set data
REFERENCE
Self-diagnostic of the GLG board is performed also at the time of equipment start.
(1) Click "CHECK DEVICE STATE" - "Check Board Status".
The Check Board Status window opens.
MU1-64
(1)
(2)
(3)
1200_400087E.ai
MU1-64