(1) Start up the MUTL.
(2) Click [SE SETTING] command, and then click [SE Registration >>] and
[SE1-10 >>].
#1 Click: [SE SETTING]
#2 Click: [SE Registration >>]
#3 Click: [SE1-10 >>]
SE SETTING window
SE1-10>> window
018-234-04E
#1
#2
SE Registration>>window
#3
1200_700098E.ai
(3) Input the serial ID of the SE (for three panel units) to be registered,
and click [SET].
u NOTE u
The first letter of the serial ID must be entered as a capital letter.
REFERENCE
The SE serial ID is contained in the folder name which is in the route for the
machine-specific CD-ROM. (Eg: L120001(TOP))
- <Serial ID for the TOP panel unit> (TOP) folder
- <Serial ID for the CENTER panel unit> (CENTER) folder
- <Serial ID for the BOTTOM panel unit> (BOTTOM) folder
#1 Input: Serial ID
#2 Click: [SET]
SE1-10>> window
(4) Check that the serial ID appears in the registration frame.
#1 Check: Serial ID
SE1-10>> window
DR-ID 1300 / DR-ID 1300PU Service Manual
#1
#2
1300_700012E.ai
<Display example>
#1
1300_700013E.ai
IN2-64