l PORT SETTING (FREE)
Specifies to which terminal block the X-ray shot cable corresponding to the operative
method [FREE (general)] of the X-ray equipment is to be connected.
<Options>
- MPX1 [default]:
Connected to MPX1
- MPX2:
Connected to MPX2
l NUM OF MP
Specifies the number of MP's to be connected.
<Settable range>
- 1 [default]:
One MP connected
- 2:
Two MP's connected
l NUM OF SE
Specifies the number of SE's (FPD's) to be connected concurrently.
<Settable range>
- 1:
One SE connected
- 2:
Two SE's connected
- 3:
Three SE's connected
- 4:
Four SE's connected
- 5 [default]:
Five SE's connected
u NOTE u
- For connecting one long panel, select "1".
- For connecting two long panels, select "2".
l TABLE DIRECTION HEAD
Selects the direction of the patient's head during exposure on the bed.
<Options>
- LEFT:
Left
- RIGHT [default]:
Right
l UPRIGHT 180 REVERSE
Selects whether the exposure image of the stand SE (FPD) is to be rotated
180 degrees.
<Options>
- none [default]:
Not rotated
- 180:
Rotated 180 degrees
018-234-03E
l READ MODE
Sets charge accumulation time of the SE (FPD). Eight options are available and are
divided into "fixed modes" and "variable modes".
- FIXED MODE
Sets constant time irrespective of the X-ray irradiation time specified in each
exposure menu of the CL. Used when the X-ray high voltage generator is not
connected.
- VARIABLE MODE
The reading mode is changed over the FIXED MODE (200 ms), the FIXED MODE
(500 ms) and the FIXED MODE (1200 ms) depending on the X-ray irradiation time
specified in each exposure menu of the CL. Used when the X-ray high voltage
generator is connected.
u INSTRUCTION u
If the X-ray irradiation time specified by the customer in the X-ray high voltage
generator is changed, the change cannot be reflected in the mode setting as the
change is not fed back to the MC before exposure. If the X-ray irradiation time
changed by the customer is longer than the charge accumulation time specified in the
MC, the irradiation is terminated upon completion of accumulation, and images with
insufficient dose may result.
Instruct the customer not to set the charge accumulation time longer than specified in
the mode when the customer is to change the X-ray irradiation time.
REFERENCE
You can check the X-ray irradiation time specified in each exposure menu through the
user utility of the CL.
DR-ID 1300 / DR-ID 1300PU Service Manual
MU2-58