(3) Wrap a plastic tape around the unconnected (remaining) cable terminals.
u NOTE u
No cable terminal may remain depending on the X-ray high voltage generator to
be connected in some cases. If no cable terminal remains, the procedure is not
necessary.
#1 Wrap: Plastic tape
#1
(4) Insert the second X-Ray shot cable (optional) from the upper opening
on the MP rear cover.
u NOTE u
If the second cable is not present, the procedure is not necessary. Proceed to the
procedure (7).
#1 Insert: X-Ray shot cable (optional)
018-234-02E
600_700014.ai
#1
1200_700030.ai
(5) Connect the cable terminals with the terminal block of the MPX54A
board.
#1 Connect: Cable terminals
MPX2
1 5V2 ISO
2 SHOT2 RDY1
3 SHOT2 RDY1
4 SHOT2 RDY2
5 SHOT2 RDY2
6 SHOT2 GO A
7 SHOT2 GO B
8 BKY2 RDY1A
9 BKY2 RDY1B
10 BKY2 RDY2A
11 BKY2 RDY2B
12 NC2
DR-ID 1300 / DR-ID 1300PU Service Manual
#1
MPX1
1 5V1 ISO
2 SHOT1 RDY1
3 SHOT1 RDY1
4 SHOT1 RDY2
5 SHOT1 RDY2
6 SHOT1 GO A
7 SHOT1 GO B
8 BKY1 RDY1A
9 BKY1 RDY1B
10 BKY1 RDY2A
11 BKY1 RDY2B
12 NC1
IN2-17
600_700016.ai
IN2-17