Guidance and manufacturer's declaration - electromagnetic immunity
The DR-ID 1300PU is intended for use in the electromagnetic environment specified below.
The customer or the user of the DR-ID 1300PU should assure that they are used in such an
environment.
Immunity test
IEC 60601 test level
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common mode
Voltage dips, short
<5% U
T
interruptions and
(>95% dip in U
voltage variations on
for 0.5 cycle
power supply input
lines
40% U
T
IEC 61000-4-11
(60% dip in U
for 5 cycles
70% U
T
(30% dip in U
for 25 cycles
<5% U
T
(>95% dip in U
for 5 s
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
018-234-02E
Electromagnetic
Compliance level
environment -
±6 kV contact
Floors should be
wood, concrete or
±8 kV air
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
±2 kV for power
Mains power quality
supply lines
should be that of a
typical commercial or
±1 kV for input/output
hospital environment.
lines
±1 kV differential
Mains power quality
mode
should be that of a
typical commercial or
±2 kV common mode
hospital environment.
<5% U
Mains power quality
T
)
(>95% dip in U
)
should be that of a
T
T
for 0.5 cycle
typical commercial or
hospital environment.
40% U
If the user of the DR-
T
)
(60% dip in U
)
ID 1300PU requires
T
T
for 5 cycles
continued operation
during power mains
70% U
interruptions, it is
T
)
(30% dip in U
)
recommended that
T
T
for 25 cycles
the DR-ID 1300PU
be powered from an
<5% U
uninterruptible power
T
)
(>95% dip in U
)
supply or a battery.
T
T
for 5 s
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Guidance and manufacturer's declaration - electromagnetic immunity
The DR-ID 1300PU is intended for use in the electromagnetic environment specified below.
The customer or the user of the DR-ID 1300PU should assure that they are used in such an
environment.
Immunity test
Conducted RF
IEC 61000-4-6
guidance
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strength from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the DR-ID 1300PU is
used exceeds the applicable RF compliance, the DR-ID 1300PU should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the DR-ID 1300PU.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
DR-ID 1300 / DR-ID 1300PU Service Manual
IEC 60601
Compliance
Electromagnetic environment - guidance
test level
level
3 Vrms
3 Vrms
Portable and mobile RF communications
150 kHz to
equipment should be used no closer to any
80 MHz
part of the DR-ID 1300PU, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1.2
3 V/m
3 V/m
d = 1.2
80 MHz to
d = 2.3
2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
level in each frequency range.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
-16
Safety Precaution
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
should be less than the compliance
b
-16
Safety Precaution