n Procedures for Connecting the Cable
CAUTION
Measure the voltage between the cables connecting to the RDY1-RDY1 and the
RDY2-RDY2 terminals before connecting the X-Ray shot cable, to make sure
that a high voltage (100 VAC, for example) is not observed whichever technique
of the X-ray high voltage generator is selected.
If the high voltage (such as 100 VAC) is observed in the cable voltage, contact a
service personnel of the X-ray equipment for checking the connection.
If erroneous connection is made, the machine might get damaged. Exercise
care.
INSTRUCTION
u
u
Always use a reusable band (clamp) included in the supplied accessories to retain the
cable, as the band is a UL-standard component. The manufacturer and the part No. of
the reusable band are mentioned below for your reference.
Manufacturer:
Kitagawa Industries, Co. Ltd.
Name:
Reusable band
Part No.:
LWS-3S V0
u NOTE u
Abut the cable terminals as shown below when fastening them together.
GOOD
Cable
Screw
Terminal
Terminal block
NO GOOD
018-234-02E
600_700132.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
(1) Insert the X-Ray shot cable (optional) and the bucky cable from the
upper opening on the MP rear cover.
#1 Insert: X-Ray shot cable (optional) and the bucky cable
X-Ray shot cable
#1
Bucky cable
1200_700034.ai
IN2-22
IN2-22