6.5
Connecting the X-Ray Shot Cable
WARNINGS
l
Be sure to turn OFF the power supply of the X-ray high voltage generator to
avoid electric shock hazards due to high voltage.
The X-ray high voltage generator is a product of another manufacturer.
l
Commit the cable connection to the service personnel of the X-ray equipment.
How to connect the X-Ray shot cable depends on the grid oscillation mode.
Refer to the corresponding connection method.
l Steady grid type
The grid is fixed and does not oscillate.
{IN1:6.5.1_Connecting the X-Ray Shot Cable (Steady Grid Type)}
l Bucky contact type
The grid is oscillated by the motor. When the relay contact is closed, the motor starts
to drive.
{IN1:6.5.2_Connecting the X-Ray Shot Cable (Bucky Contact Type)}
l Bucky AC type
The grid is pressed against the leaf spring, and the grid rebounds by means of
reaction force of the spring. A solenoid is employed as a mechanical part against
which the grid is pressed.
The AC bucky relay unit (optional) is required when the bucky AC type is to be
connected.
{IN1:6.5.3_Connecting the X-Ray Shot Cable (Bucky AC Type)}
018-234-04E
6.5.1
Connecting the X-Ray Shot Cable (Steady Grid Type)
n Connection Diagram of the X-Ray High Voltage Generator (When
a Single SE Is Connected)
X-ray high
voltage generator
SHOT1
RDY 1
Shot
Switch 1
SHOT1
Prep
RDY 1
SHOT1
RDY 2
Shot
Switch 2
SHOT1
RAD
RDY 2
SHOT1
GO A
SHOT1
GO B
DR-ID 1300 / DR-ID 1300PU Service Manual
MPX54A
+5V
K5
AHN22005
+5V_iso
+5V_iso
Driver
circuit
1
PC3
+5V
TLP421F
2
820Ω
K6
AHN22005
+5V_iso
3
PC2
TLP421F
GND_iso
4
820Ω
5
GND_iso
+5V
6
K4
AHN22005
7
IN1-14
BKY1
8
RDY1A
BKY1
9
RDY1B
BKY1
10
RDY2A
BKY1
11
RDY2B
12
600_700010.ai
IN1-14

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dr-id 1300pu

Table of Contents