Error
Error Name
Code
12902
System configuration error
12910
Control timeout detected
Unknown machine connection
12920
error
Software logic error (only in
12990
log)
13043
Electrical noise detection error
018-234-06E
Occurrence Condition
[During bootup/routine processing]
An item that needs to be edited from its default value is not changed in
EDIT CONFIG.
Example) MC ID Code
[During bootup/routine processing]
The application software detects control timeout.
[During bootup]
A machine to which the standard sequence cannot pass is connected
during connection.
[During bootup/routine processing]
The application program detects a logic error for software control.
[During routine processing]
A noise error is detected in the panel.
DR-ID 1300 / DR-ID 1300PU Service Manual
Probable Cause and Remedy
- There are missing settings in the config.
- Check communication among the MC, the MP and the
FPD.
- Check the connection with the AP (for wireless type).
- Check the connection of the SE cables (for wired type).
- Added from MC V16.3. The ready lamp turns off when
an electrical noise error occurs, and an error message
is displayed after the exposure cannot be done.
- Click [OK] and check that the ready lamp turns on again.
<NOTE>
When a 13043 error occurs, 11112 (detail code:
0x00800000) is recorded in the error log.
MT-18