Relation among configurations
X-CON
MP1
SE1
SE serial No.
MPC7
MPC6
SE2
SE serial No.
Point 2: SE serial No.
Serial Nos. of all panels are registered in advance via the PC-TOOL.
The operative method for exposure and the panel to be used are determined based on the
CSL setting (not depending on the connector (MPC6 or 7) to which the panel is connected).
018-234-03E
Point 1: X-ray shot cable
The name of a connector to which a cable of each operative method
is connected is registered in advance via the PC-TOOL.
Free
Table
[EDIT CONFIGURATION]
Upright
[OP SETTING(FREE)]
[OP SETTING(TABLE)]
[OP SETTING(UPRIGHT)]
[MP1] or
[MP2] or
[None]
[PORT SETTING(UPRIGHT)]
MPX1
[MPX1] or
MPX2
[MPX2] or
[None]
[MUTL] - [SE Registration]
SE1: SE serial No.
SE2: SE serial No.
SE3: SE serial No.
SE4: SE serial No.
SE5: SE serial No.
MC
[Device Name]: Panel identification name
[FPD Device ID]: SE Serial No.
[Device Type]:
[Device Color]: Identification of SE
DR-ID 1300 / DR-ID 1300PU Service Manual
CSL
[Selector Setting]
[Selector12]
[Selector1]
Defines the operative method.
[Technique Code]:
[Reader Type]:
[FPD-C]: FREE exposure with FPD or
[FPD-T]: TABLE exposure with FPD or
[FPD-U]: UPRIGHT exposure with FPD
[Caption]: Selector name
[Icon Type]: Selector icon
[Device Name]: Panel identification name
[Device Setting]
Defines the panel for use.
[D-EVO] or
[AcSelarate]
Image processing is selected
depending on the type of the panel.
MU2-66
1300_400004E.ai
MU2-66