(3) Connect the cable terminals to the terminal blocks (TB1 and TB2) of
the AC bucky relay unit.
WARNING
Be sure to connect the X-Ray shot cable and the FG line of the bucky cable to
the FG (TB2-6) of the terminal block. If the FG line is fixed to the MP frame or the
like, electrification or ignition might result due to difference in ground levels.
u NOTE u
If two X-Ray shot cables and two bucky cables are to be connected, make
connection of the first cable to the terminal indicated in a white cell and the
second cable to the terminal indicated in a shaded cell shown in the table below.
The method of connecting the second cable is described in the procedure (4).
No.
TB1
X-Ray shot cable
Bucky cable
1
X-Con BKY
SHOT1 RDY1 (L)
2
X-Con BKY
SHOT1 RDY1 (N)
3
X-Con
SHOT1 RDY2 (L)
4
X-Con BKY
SHOT1 RDY2 (N)
5
BKY
SHOT1 RDY2 (L)
6
X-Con BKY
SHOT2 RDY1 (L)
7
X-Con BKY
SHOT2 RDY1 (N)
8
X-Con
SHOT2 RDY2 (L)
9
X-Con BKY
SHOT2 RDY2 (N)
10
BKY
SHOT2 RDY2 (L)
018-234-02E
TB2
X-Ray shot cable
Bucky cable
X-Con BKY
Stand AC (L)
X-Con BKY
Stand AC (N)
X-Con BKY
Bed AC (L)
X-Con BKY
Bed AC (N)
N.C.
N.C.
X-Con BKY
Stand FG
X-Con BKY
Bed FG
X-Con BKY
Stand Exp A
X-Con BKY
Stand Exp B
X-Con BKY
Bed Exp A
X-Con BKY
Bed Exp B
TB1
1
2
3
4
5
6
7
8
9
10
DR-ID 1300 / DR-ID 1300PU Service Manual
X-Ray shot cable
TB2
Bucky cable
1
2
3
4
5
6
7
8
9
10
600_700050.ai
IN2-32