FujiFilm DR-ID 1300 Service Manual page 402

Digital radiography
Table of Contents

Advertisement

n Signal Descriptions
l SHOT1 RDY1/RDY1 (Input)
The signal is generated when only the first stage of the shot switch on the X-ray high
voltage generator is pressed. (Prep signal)
l SHOT1 RDY2/RDY2 (Input)
The signal is generated when the first and second stages of the shot switch on the
X-ray high voltage generator are pressed. (RAD signal)
l BKY1 RDY1 A/B (Output)
The signal prepares for the bucky operation.
A relay type is normally open (NO).
l BKY1 RDY2 A/B (Output)
The signal operates the bucky upon completion of the preparation for the SE
exposure.
A relay type is normally open (NO).
n Contact Capacity
The specifications of relays (K4, K5 and K6) are as follows.
l Relay model
AHN22005 (Panasonic Electric Works)
l Rated control capacity
Standard contact: 30 VDC
l Maximum switching current
1 A
018-234-04E
n Timing Chart
A: The signal for preparing for the bucky operation is output, and preparation starts for SE
exposures corresponding to the first stage of the switch when the shot switch (first stage)
is turned ON.
B: Preparation starts for SE exposures corresponding to the second stage of the switch
when the shot switch (second stage) is turned ON.
C: The signal for operating the bucky is output upon completion of the preparation for the SE
(FPD) exposure.
D: An exposure request signal is output after a predetermined time elapses since the bucky
operation starts, and X-ray exposure starts.
E: X-ray exposure ends.
F: Charge accumulation of the SE (FPD) ends.
SHOT1 RDY1/RDY1
BKY1 RDY1/RDY1
SHOT1 RDY2/RDY2
BKY1 RDY2/RDY2
Bucky
BKY1 EXP
X-ray exposure
Charge accumulation
of SE (FPD)
DR-ID 1300 / DR-ID 1300PU Service Manual
A
B C
D
E F
600_700023.ai
IN1-21
IN1-21