FujiFilm DR-ID 1300 Service Manual page 180

Digital radiography
Table of Contents

Advertisement

 MODE 2 EXPOSURE TIME
Sets the charge accumulation time of READ MODE 2. Set in a range of 1.2 to 3.8 s.
The setting can be changed in 0.2-s steps.
<Settable range>
- 1.2 to 3.8 s [default: 1.2 s]
Charge accumulation time of READ MODE 2
u NOTE u
For the facilities that may use the machine long time under high temperature, to
prevent artifact, it is recommended to set the charge accumulation time of READ
MODE 2 at or lower than 1.2 seconds.
 MODE 3 EXPOSURE TIME
u NOTE u
Do not use this item.
 Smart Switch HIGH SENSITIVE MODE TARGET
u NOTE u
Do not use this item.
AUTO DETECT
Set whether to use the automatic X-ray detection function (AUTO DETECT) or not.
<Options>
- OFF [default]:
Not used
- ON:
Used
 CONNECTING CU
Set to CONNECTED when the machine is to be connected to the DX Console to
which AcSelarete or CALNEO U/T is connected.
Usually, use the setting of NOT CONNECTED [default].
<Options>
- NOT CONNECTED [default]
- CONNECTED
018-234-04E
 EMERGENCY MODE
Set whether the emergency mode is to be used or not.
<Options>
- OFF [default]:
- ON:
REFERENCE
Emergency mode is the mode that the calibration would be stopped automatically and
the exposure would be ready when the system starts.
When it is set to ON, the exposure would be ready as soon as the system starts.
 BATTERY WARNING[min]
Sets the threshold value of remaining battery charge.
If the remaining battery charge lowers to the set threshold value, exposure cannot be
accepted. The SE LED flashes in this case.
The setting is made as the remaining battery time (minutes).
Always use the setting of 10 [default] in this machine.
<Settable range>
- 0 to 120 minutes [default: 10]
 SE BATTERY WATCHING (when connecting cables)
Set this option to ON for a configuration, such as the mobile configuration, wired with
the SE but the battery of which is not charged.
Set this to OFF [default] for a configuration including the MP, DS, and PB.
<Options>
- OFF [default]:
- ON:
REFERENCE
This function depends on whether this machine supplies power to the SE.
- Supply power to and communicate with the SE.
Irrespective of ON or OFF, the battery is monitored.
- Supply no power to but communicate with the SE.
If setting to "OFF", does not activate monitoring of the battery.
If setting to "ON", activates monitoring the battery.
DR-ID 1300 / DR-ID 1300PU Service Manual
Not used
Used
Not watched
Watched
MU1-15