FujiFilm DR-ID 1300 Service Manual page 178

Digital radiography
Table of Contents

Advertisement

 PORT SETTING (TABLE)
Specifies to which terminal block the X-ray shot cable corresponding to the operative
method [Bed] of the X-ray equipment is to be connected.
<Options>
- MPX1 [default]:
Connected to MPX1
- MPX2:
Connected to MPX2
 PORT SETTING (FREE)
Specifies to which terminal block the X-ray shot cable corresponding to the operative
method [FREE (general)] of the X-ray equipment is to be connected.
<Options>
- MPX1 [default]:
Connected to MPX1
- MPX2:
Connected to MPX2
 NUM OF MP (s)
Specifies the number of MPs to be connected.
u NOTE u
Specify only the number of MPs. Do not include the number of PBs and DSs.
<Settable range>
- 0 [default]:
Not connected
(such as a mobile configuration, a configuration using the DS only, etc.)
- 1:
One MP connected
- 2:
Two MPs connected
 NUM OF SE (s)
Specifies the number of SEs (FPDs) to be connected concurrently.
<Settable range>
- 1:
One SE connected
- 2:
Two SEs connected
- 3:
Three SEs connected
- 4:
Four SEs connected
- 5 [default]:
Five SEs connected
u NOTE u
- For connecting one long panel, select "1".
- For connecting two long panels, select "2".
018-234-04E
 TABLE DIRECTION HEAD (for DR-ID 1422SE)
Selects the direction of the patient's head during exposure on the bed.
<Options>
- LEFT:
- RIGHT [default]:
 UPRIGHT 180 REVERSE
Selects whether the exposure image of the stand SE (FPD) is to be rotated
180 degrees.
<Options>
- none [default]:
- 180:
DR-ID 1300 / DR-ID 1300PU Service Manual
Left
Right
Not rotated
Rotated 180 degrees
MU1-13
MU1-13