FujiFilm DR-ID 1300 Service Manual page 92

Digital radiography
Table of Contents

Advertisement

Error
Error Name
Code
Insufficient number of
12235
exposures for residual image
calibration
Defect map compression/
12236
decompression failure
Exposure menu error (SE not
12240
connected)
Exposure menu error (serial ID
12242
not registered)
12243
Image transfer error (log level)
Exposure menu error (panel
12245
degeneration)
Panel communication error (log
12246
level)
12250
DR cassette unit error
12255
Long panel error
Marker calibration data
12280
generation error
Insufficient number of
12281
exposures for marker
calibration
018-234-06E
Occurrence Condition
[During full calibration]
After residual image calibration is to take place, the number of exposures
necessary for the residual image calibration is recognized to be
insufficient.
<Reference>
A single exposure is necessary to carry out the residual image calibration.
Compression or decompression of the defect map fails.
[During exposure menu registration]
The exposure menu cannot be accepted because the SE is not
connected.
[During exposure menu registration]
The exposure menu cannot be accepted because the SE serial ID is not
registered.
[During routine processing]
Image transfer fails. Or an error occurs in the exposure sequence.
[During exposure menu registration]
The exposure menu cannot be accepted because the panel is
degenerated.
[During exposure menu registration]
An error occurs in a control command to the panel.
[During bootup/routine processing]
An error is detected in the DR cassette unit.
[During bootup/routine processing]
An error is detected in the long panel hardware or the SE setting.
[During full calibration]
An error is detected in marker calibration data generation.
[During full calibration]
After marker calibration is to take place, the number of exposures
necessary for the marker calibration is recognized to be insufficient.
DR-ID 1300 / DR-ID 1300PU Service Manual
Probable Cause and Remedy
- Check the exposure count.
- Continue exposures until the necessary exposure count
for the residual image calibration is reached, and click
[Create a Gain Data].
- Check communication among the MC, the MP and the
FPD.
- Check the connection with the AP (for wireless type).
- Check the connection of the SE cables (for wired type).
- Register the serial ID of the SE.
- Check communication among the MC, the MP and the
FPD.
- Check the connection with the AP (for wireless type).
- Check the connection of the SE cables (for wired type).
- Select a defective item from the detail error log, and
check it.
- Check communication among the MC, the MP and the
FPD.
- Check the connection with the AP (for wireless type).
- Check the connection of the SE cables (for wired type).
- Check the long panel setting.
- Check the marker position and exposure conditions.
- Check the hardware error.
- Check whether the exposure count has turned to the
fourth exposure.
MT-15
MT-15

Hide quick links: