FujiFilm DR-ID 1300 Service Manual page 397

Digital radiography
Table of Contents

Advertisement

n Procedures for Connecting the Cable
CAUTION
Measure the voltage between the cables connecting to the RDY1-RDY1 and the
RDY2-RDY2 terminals before connecting the X-Ray shot cable, to make sure
that a high voltage (100 VAC, for example) is not observed whichever technique
of the X-ray high voltage generator is selected.
If the high voltage (such as 100 VAC ) is observed in the cable voltage, contact a
service personnel of the X-ray equipment for checking the connection.
If erroneous connection is made, the machine might get damaged. Exercise care.
INSTRUCTION
u
u
Always use a reusable band (clamp) included in the supplied accessories to retain the
cable, as the band is a UL-standard component. The manufacturer and the part No. of
the reusable band are mentioned below for your reference.
Manufacturer:
Kitagawa Industries, Co. Ltd.
Name:
Reusable band
Part No.:
LWS-3S V0
(1) Insert the X-Ray shot cable (optional) from the upper opening on the
MP rear cover.
#1 Insert: X-Ray shot cable (optional)
018-234-04E
#1
1200_700029.ai
(2) Connect the cable terminals with the terminal block of the MPX54A
board.
#1 Connect: Cable terminals
MPX2
1 5V2 ISO
2 SHOT2 RDY1
3 SHOT2 RDY1
4 SHOT2 RDY2
5 SHOT2 RDY2
6 SHOT2 GO A
7 SHOT2 GO B
8 BKY2 RDY1A
9 BKY2 RDY1B
10 BKY2 RDY2A
11 BKY2 RDY2B
12 NC2
DR-ID 1300 / DR-ID 1300PU Service Manual
MPX1
1 5V1 ISO
#1
2 SHOT1 RDY1
3 SHOT1 RDY1
4 SHOT1 RDY2
5 SHOT1 RDY2
6 SHOT1 GO A
7 SHOT1 GO B
8 BKY1 RDY1A
9 BKY1 RDY1B
10 BKY1 RDY2A
11 BKY1 RDY2B
12 NC1
IN1-16
600_700013.ai
IN1-16

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dr-id 1300pu

Table of Contents