FujiFilm DR-ID 1300 Service Manual page 298

Digital radiography
Table of Contents

Advertisement

[5.1.1] MP1 Board State >>
 Display Window
MP1 Board State>>window
018-234-04E
1200_400025E.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
 Function Description
Acquires the conditions of components connected with the board in the MP1. You can
view the acquired results in the result display area.
- Connector Status:
The status of the connector connected with the MP board appears in the result
display area.
Normal processing: MPX Connector: ON
Upright SE Connector: ON
(when connected with the MPC board's MPC6 connector cable)
Table SE Connector: ON
(when connected with the MPC board's MPC7 connector cable)
Error:
Other than above.
- Fan Status:
The status of the fan connected with the MP board appears in the result display
area.
Normal processing: MP Fan Status: ON
Error:
Other than above.
- Fuse Status:
The status of the fuse connected with the MP board appears in the result display
area.
Normal processing: MPX Fuse: OK, UprightSE Fuse: OK, TableSE Fuse: OK
Error:
Other than above.
- RemoteSW Status:
The ON/OFF status of the remote switch appears in the result display area.
ON: OFF/[ON]
OFF: [OFF]/ON
- ShotSW Port1 Status
The ON/OFF status of the shot switch port 1 appears in the result display area.
ShotSW1 is ON:
OFF/[ON]
ShotSW1 is OFF: [OFF]/ON
- ShotSW Port2 Status
The ON/OFF status of the shot switch port 2 appears in the result display area.
ShotSW2 is ON:
OFF/[ON]
ShotSW2 is OFF: [OFF]/ON
MU2-33
MU2-33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dr-id 1300pu

Table of Contents