FujiFilm DR-ID 1300 Service Manual page 562

Digital radiography
Table of Contents

Advertisement

n Changing the Calibration Position and Exposing Calibration
(1) Change the setting of [Calib Position] from the value selected in
the step (6) of "■ Preparing for Calibration" to "Center_Bottom".
Then, perform "■ Preparing for Calibration", "■ Installing Calibration
Phantom", "■ Exposing Calibration", and "■ Creating Calibration Data
and Checking Result".
018-234-04E
■ Checking the Image Composition Accuracy
(1) Left-click the MC Manager from the task tray and execute "EXIT".
Check that the MC Manager icon has disappeared from the task tray, and then
start the MC Manager from "Start menu" → "Start-up".
(2) Create the confirmation menu.
Start the User Utility and copy the "Finger" menu (MPM code: 0607), and then
edit the following items as shown below. With the other settings left unchanged,
exit the User Utility.
- Menu name: Synthesized calibration checking test
- Function: Leave the setting as "RT"
(3) Start up the application for the DX Console.
(4) Select the check menu created in step (2), and perform exposure
at the two positions shown below with the calibration phantom left
placed. Visually check the connection of the calibration phantom for
misalignment.
- Tube voltage of 80 kV, Amperage: 50 mA, Time: 13 msec, SID: 180 cm
1st shot
u NOTE u
If a misalignment is found, perform the full calibration and marker calibration for
the CENTER panel unit again.
DR-ID 1300 / DR-ID 1300PU Service Manual
2nd shot
1300_700036E.ai
IN2-85
IN2-85