FujiFilm DR-ID 1300 Service Manual page 88

Digital radiography
Table of Contents

Advertisement

Error
Error Name
Code
Wrong irradiation detection
11229
error
Invalid data generated for
11231
defect correction
11232
Panel noise warning
11239
Initialization calibration error
11240
Panel disconnection detected
11242
Unregistered panel detected
Image acquisition timeout
11243
(between panel and MC)
SE communication error
11246
(warning level)
018-234-06E
Occurrence Condition
X-ray irradiation is performed before the panel turns on the Ready state.
[During bootup]
The value of the maximum defect pixel size or defect pixel density
exceeding the specification is detected in defect correction data
generation of background calibration (calibration automatically taking
place when the CU starts up).
[ During full calibration]
The value of the maximum defect pixel size or defect pixel density
exceeding the specification is detected in defect correction data
generation of full calibration.
<Reference>
You can know in which specifications the error is detected on the error
log.
[During routine processing]
The SE has detected electrical noise.
[During initialization calibration]
The calibration requirement for the SE is NG.
[During bootup/routine processing]
Disconnection of the line to the panel is detected.
[During bootup/routine processing]
A panel with an unregistered serial ID is detected.
[During routine processing]
An error occurs in the image acquisition sequence from the panel.
[During bootup/routine processing]
An error occurs in a control command to the panel.
DR-ID 1300 / DR-ID 1300PU Service Manual
Probable Cause and Remedy
- Urge users to pay attention to performing the exposure
after checking the Ready state because the X-ray
irradiation is performed before the panel turns on the
Ready state.
- Check the FPD.
- Restart the panel.
- If this occurs multiple times, then check the noise
source.
- This might occur if X-rays were irradiated during the
initialization calibration, or if it was affected by magnetic
field noise from nearby electronic devices which were
being operated nearby. If this occurs several times, then
check reducing the noise.
In addition, even if the initialization calibration fails, if
the subsequent offset calibration succeeds, and no
abnormalities are generated in the exposure images,
then there is no problem.
- Check the switch-HUB incorporated in the MP.
- Check the network interface board incorporated in the MC.
- Check the wireless module.
- Check the POW board.
- Check the RMV board.
- Check the cable connecting between the POW board
connector and the GTD board connector.
- Check the wireless module.
- Check the POW board.
- Check the RMV board.
- Check communication between the FPD and the MP.
- Check the connection with the AP (for wireless type).
- Check the connection of the SE cables (for wired type).
MT-11
MT-11