FujiFilm DR-ID 1300 Service Manual page 210

Digital radiography
Table of Contents

Advertisement

 Read images
Since the read images information is displayed in the upper left frame of the
window, check that the pixel values and Image Type are correct.
Example) 17x17 sized image: 2832 x 2836 Pixels, 16-bit
POCX01020D009.ai
No good example)
If the Image Type is incorrect
Read at 8 bits
018-234-06E
No good example)
If the SE size pixel values are incorrectly input
17x17 is read at 14x17
POCX01020D010E.ai
(4) Select the Image – Zoom – View 100% menu, and check the line defect
locations on the pixel same size images.
REFERENCE
Change the tones in the following menu as need be, to make them easily visible.
Image – Adjust – Brightness/Contrast
By using the "hand" icon, the image display positions can be changed.
(5) If line defects are found, press the magnify button in the following
figure to enlarge and display the line defects up to 600%, and check
the line defect coordinate positions.
REFERENCE
The coordinates will be registered if the vertical line defects are x coordinates,
and if the horizontal line defects are y coordinates.
DR-ID 1300 / DR-ID 1300PU Service Manual
MU1-45
1200_400171.ai
MU1-45