FujiFilm DR-ID 1300 Service Manual page 449

Digital radiography
Table of Contents

Advertisement

NOTE
- When the CPU software installed in the SE is old
If the CPU software installation fails during a software update, check the CPU
software version.
If the CPU software version is less than 0x0220 (corresponding MC V11.0),
this function cannot be updated because the CPU software installed in the SE
is old.
Upgrade the SE application version again after updating the FPGA software
from "OPERATION>UTILITY> Update SE CPU (0x220)".
{MU1:1.9.10_Update SE CPU (0x0220)}
- When the FPGA software installed in the SE is old
When updating the software, if the message "This function is not supported by
the selected SE" is displayed and the FPGA software installation fails, check
the version of the FPGA software.
If the FPGA software version is earlier than the following version, the FPGA
software installed in the SE is old and cannot be updated directly.
Panel
fpgaRMV (FPGA) software
DR-ID 1201SE
0x0080
DR-ID 1202SE
0x00A0
DR-ID 1211SE
0x00B0
DR-ID 1212SE
0x00B0
DR-ID 1213SE
0x00B8
DR-ID 1214SE
0x00C8
DR-ID 1305SE
0x00A0
Upgrade the SE application version again after updating the FPGA software
from "OPERATION>UTILITY> Update SE FPGA".
{MU1:1.9.11_Update SE FPGA}
018-234-06E
(2) Check that the installation is completed, click [OK], and restart the SE.
(3) Click [GET], confirm that the SE application version has been updated,
and click [Next].
{n Checking the SE Application Software Version}
NOTE
With the DR-ID 1300, do not perform "Setting the SE LED Color".
DR-ID 1300 / DR-ID 1300PU Service Manual
1200_700507E.ai
1200_700508E.ai
IN1-68
IN1-68

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dr-id 1300pu

Table of Contents