(1) Select the target operative method from the drop-down list box.
(2) Click [START].
u NOTE u
Clicking [START] during automatic offset update (approximately 30 seconds
every 10 minutes), "Error 12700 currently unavailable" is displayed. Again click
[START] approx. 30 seconds later in this case.
REFERENCE
Whether the calibration is successfully completed appears after completion.
An error code is displayed in some cases. Refer to the Troubleshooting manual
and take appropriate countermeasures.
After about 2 minutes, the [Ready] button will be displayed in green in the pop up
window.
u NOTE u
If you perform exposure before [Ready] is lit up with the green button, calibration
will fail and create an abnormal image.
(3) Expose 1 time with the following condition.
1305SE: Tube voltage of 80 kV, dose of 100 mR
Images for lag correction data generation (7 frames) are read from the SE (FPD).
Images of seven frames are read with a predetermined interval for one exposure.
The number of image reads is displayed in the format of "n/7".
u INSTRUCTION u
Set the exposure time to 200 msec or less. Exposure for longer than 200 msec
cannot be made, since the maximum exposure time is inherently specified as
200 msec.
018-234-06E
u NOTE u
- In the beginning when afterimage calibration is implemented, record the
exposing conditions, and subsequently be sure to expose with the same
conditions.
This is for comparing the afterimage amounts over time, so that SE (FPD)
degradation analysis can be done in advance.
- If 100mR or greater irradiating cannot be performed via conditions with a
80kV tube voltage, and maximum irradiating times of 200 ms, then close the
distance between the X-ray tube and the panel as much as possible (to the
limit whereas the entire panel will be irradiated by the X-rays).
Nevertheless, if 100mR or greater irradiating cannot be performed, raise the
tube voltage up to 100kV. With these conditions, if the irradiating is 20mR or
greater, implement the afterimage correction calibrating.
If the aforementioned conditions do not reach 20mR or greater, the afterimage
correction calibrating needs to be implemented via another X-ray device.
When the log correction data is automatically generated and stored in the HDD,
"Succeeded" appears after completion.
u NOTE u
If "NG" appeared, check the following.
- Exposure condition
- Exposure field is the entire SE.
(4) Click [OK].