● Difference in image density between panels (second example): MC
V12.0 or later
Precautions for the Virtual Grid:
The image density may change at the stitching part of the panels. Especially when the
Virtual Grid processing is applied to a thick object, the difference in density is more
clearly visible.
Anterior lumbar + Acrylic 5 cm (no grid)
Position of
unevenness
at stitching
part
Virtual Grid processing
Position of
unevenness
at stitching
part
018-234-03E
1300_700045E.ai
● Difference in image density between panels (third example)
The difference in the image density may be visible not entirely but partly in the
horizontal direction along the stitching area.
Position of
unevenness
at stitching
part
● White band and black band (first example)
Especially for a thick object, band-like artifacts may appear.
DR-ID 1300 / DR-ID 1300PU Service Manual
No difference
in density is visible.
Little whitish
Little blackish
1300_700044E.ai
Position of
unevenness
at stitching
part
Position of
unevenness
at stitching
part
AppxIN4-3
Black band
White band
1300_700046E.ai
AppxIN4-3