FujiFilm DR-ID 1300 Service Manual page 85

Digital radiography
Table of Contents

Advertisement

Error
Error Name
Code
10750
MPX fuse blown error
10751
MP fan stop
10752
MPX connector disconnected
10753
MP connection error
10900
Parameter error
10902
System configuration error
10990
Software logic error
10991
Parameter file key not hit
11111
RMV board blown fuse detected
018-234-06E
Occurrence Condition
[During bootup/routine processing]
A blown fuse of the MPX board is detected, and the system shuts down.
[During bootup/routine processing]
MP cooling fan (FAN1) stop is detected, and the system shuts down.
[During bootup/routine processing]
The system shuts down, as the cable is disconnected from the MPX
board or the cable is unconnected.
[During bootup/routine processing]
Disconnection of a line between the MP and the MC is detected.
[During bootup]
Parameter settings for the stand and the bed are not made, and the
system shuts down.
[During bootup]
The following mismatches are found when the config is checked during
startup of the MC:
- Emergency mode ON & Long-time accumulation ON
- AutoDetect OFF & Long-time accumulation ON
- Third panel configuration ON & Emergency mode OFF
[During bootup/routine processing]
An error is detected in the logic for controlling the software, and the
system shuts down.
[During bootup/routine processing]
An attempt is made to load the parameter file, but the key is not detected,
resulting in system shutdown.
[During bootup/routine processing]
- The RMV board detects blowout of a fuse of the POW board mounted
on the FPD, and disables the FPD.
- The RMV board detects a failure in connection of a power connector of
the GTD board mounted on the FPD, and disables the FPD.
- The RMV board detects an error of the MP cooling fan (P-FAN), and
disables the FPD.
[During M-Utility]
Blowout of a fuse of the POW board is detected by the RMV board during
the check of fuses.
<Reference>
You can know which device (fuse, connector or fan) is defective on the
error log.
DR-ID 1300 / DR-ID 1300PU Service Manual
Probable Cause and Remedy
- Check the fan (FAN1).
- Check the MPC board.
- Check the MPX board.
- Check the MP power. (If the power of the MP is OFF,
turn it ON.)
- Select a defective item from the detail error log
mentioned in ERRORDB.LOG, and check it.
- Check the parameters.
- An invalid combination is set in the config.
- Check the FPD.
- Check the MP cooling fan (P-FAN).
MT-8
MT-8