Appendix 9. Installing FPGA onto
the New FPGA Memory
Equipped SE
In SEs equipped with new FPGA memories, the FPGA software cannot be installed
from MCs with MC V17.1 or earlier.
Therefore, the jig software (FlashUpdate.zip) which can install the FPGA into older
MC compatible versions is provided for setting up the target SEs in older MC software
environments.
n Target OS
Windows7 and Windows10
n FPGA version of target SE
Panel
FPGA Software
DR-ID 1201SE
0x0085 or later
DR-ID 1202SE
0x00A5 or later
DR-ID 1211SE
0x0095 or later
DR-ID 1212SE
0x00B5 or later
DR-ID 1213SE
0x00BD or later
DR-ID 1214SE
0x00CD or later
DR-ID 1305SE
0x00A5 or later
n Target part
Service part numbers for RMV equipped with the new FPGA memory are shown in the
table below.
Part
Current part number
RMV65A service part for
857Y120057
DR-ID 1200
RMV66A service part for
857Y120105
DR-ID 1200
RMV65A service part for
857Y120005
DR-ID 1300
018-234-06E
New part number
(New FPGA memory equipped part)
857Y200090
857Y200091
857Y200089
n Confirm the resolution details
l Confirm the timing
- When the target SE is connected to the MC
- When the SE's internal board is replaced with a board in the target service parts
l Confirm the MC software version
Check the installed MC software version. Then, perform the appropriate action
explained in the table below.
MC Software
V17.1 or later
No action is needed.
V16.7
Upgrade the MC to V17.1 or later. If version upgrade is impossible, no
action is needed.
V15.1 to V16.6
Upgrade the MC to V17.1 or later. If version upgrade is impossible,
install the FPGA by using the jig software.
V15.1 or earlier
Upgrade the MC to V17.1 or later.
The jig software cannot be used for MC earlier than V15.1.