FujiFilm DR-ID 1300 Service Manual page 192

Digital radiography
Table of Contents

Advertisement

 Gain Calibration
Carries out gain calibration.
- Exposure count: 16
018-234-06E
(1)
(2)
1200_400070E.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
(1) Select the target operative method from the drop-down list box.
(2) Click [START].
If the exposing preparations are finished, then [Ready] will be displayed on the
green button in the pop up window.
u NOTE u
Clicking [START] during automatic offset update (approximately 30 seconds
every 10 minutes), "Error 12700 currently unavailable" is displayed. Again click
[START] approx. 30 seconds later in this case.
REFERENCE
Whether the calibration is successfully completed appears after completion.
An error code is displayed in some cases. Refer to the Troubleshooting manual
and take appropriate countermeasures.
u NOTE u
Before going to the next step (exposure), check that [Ready] is lit up with the
green button.
If you perform exposure before [Ready] is lit up with the green button, calibration
will fail and create an abnormal image.
(3) Expose 16 times with the following condition. Confirm that [Ready]
lights with the green button at each exposure.
u NOTE u
Perform the next exposure every time after [Ready] is lit up with the green button
because the [Ready] disappears every exposure.
1305SE: Tube voltage of 75 kV, dose of 10 mR
The gain correction data generation images (16 frames) are read out from the SE
(FPD).
u INSTRUCTION u
Set the exposure time to 200 msec or less. Exposure for longer than 200 msec
cannot be made, since the maximum exposure time is inherently specified as
200 msec.