n Procedures for Replacing the MP or the MPC54B Board
(1) Replace the MP or the MPC54B board.
REFERENCES
Refer to the following for replacing the MPC54B board.
{MC:2.3_MPC54B Board}
(2) Update MP application software version.
{IN1:10.5_Updating MP Application Software Version}
{IN2:10.5_Updating MP Application Software Version}
REFERENCES
By changing the board, the IP address has been changed to 192.162.0.20 (default
value) which is equivalent to MP1. Therefore, update MP application software
version by clicking [MP1 Software VerUp Install] on the MP1 Install >> window.
n Changing the Local IP Address
(1) Return the local IP address of the SE/MP/MC to the original value (value
before changed to the default value).
{IN:Appendix 3._Replacement Procedure of the Local Network}
REFERENCE
It is not necessary to change the IP address of the SE.
(2) Restore the respective data which has been backed up.
- CONFIGURATION
- HISTORY LOG
- CORRECT ALL DATA
{MU2:1.14_RESTORE}