FujiFilm DR-ID 1300 Service Manual page 560

Digital radiography
Table of Contents

Advertisement

n Exposing Calibration
(1) Perform positioning under the following conditions:
- Move the irradiation field center of the tube to the position that matches the
center (red cross) of the overlapping part of the panels.
- Make the distance 120 cm from the panel surface to the focal spot of the tube.
- Maximize the irradiation field of the collimator.
- Move the tube in the short direction of the panel up to the position where the
right end of the light irradiation field overlaps with the inside of the panel frame.
(2) Expose the 1st shot with the following conditions. Click [Display
Image Count] and check that exposures were counted.
- Tube voltage of 80 kV, Amperage: 100 mA, Time: 20 msec, SID: 120 cm
Marker Calibration >> window
018-234-04E
1300_700031.ai
1300_700032E.ai
(3) Move the irradiation field in the opposite direction as shown below.
1300_700033.ai
(4) After clicking [Display ShotRdy] and confirming the display of "SHOT_
READY" in the result indication column, expose the 2nd shot with
the following conditions. Click [Display Image Count] and check that
exposures were counted.
- Tube voltage of 80 kV, Amperage: 100 mA, Time: 20 msec, SID: 120 cm
(5) After clicking [Display ShotRdy] and confirming the display of "SHOT_
READY" in the result indication column, expose the 3rd shot with
the following conditions. Click [Display Image Count] and check that
exposures were counted.
- Tube voltage of 80 kV, Amperage: 100 mA, Time: 20 msec, SID: 120 cm
u NOTE u
Do not move the irradiation field.
1300_700033.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
IN2-83
IN2-83